»  Categories   »      Temperature Forcing System   »    Environmental   »    Thermostream  

image enlarge

Temptronic ATS-810-M Thermostream -80° to +225°C

Description

Designed for 60Hz operation only, this Advanced Temperature Source for fast and precise thermal conditioning of components, parts, hybrids, modules, sub-assemblies, and printed circuit boards. Capable of ultra-low temperatures without the use of Liquid Nitrogen (LN2) or Liquid Carbon Dioxide (LCO2).

Features

  • Transition Rates:
              -55°C to +125°C, <10 seconds
              +125°C to -55°C, <10 seconds
  • System Airflow Output: 4 to 18 scfm ( 1.8 to 8.5  l/s) Continuous
  • Temperature  Display & Resolution: ±1°C
  • Temperature Accuracy: ±1°C
  • Directly Control the device under test (DUT)
  • DUT Control Sensors: Type T or Type K Thermocouples, RTD (100 Ohm)
  • Remote interfaces: IEEE.488, RS232
  • Thermal Cap provides a localized thermal test environment
       protects from moisture and condensation at the DUT site
  • LabView™ drivers
  • CE Compliant and CFC-free
  • Heated Defrost to quickly remove moisture buildup from the internal chiller
  • Heat Only Mode to reduce power usage when cold temperatures are not used
  • Automatic Power Reduction to reduce power usage during idle periods

Specifications

Facility Requirements

Power

60Hz only, system does not operate at 50Hz

200 - 250 VAC (230V nominal), 60Hz, 30 amp, 1phase

Compressed Air

Clean, Dry Air (CDA)

Filtered to 5 micron particulate contamination.

Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.

Dew point: <-10°C @ 6.2 BAR (90PSI)

Air Supply Pressure

6.2  to 7.6 BAR (90 to 110 PSIG)

Total Air Flow Rate Required

7.1 to 14.2 l/s (15-30 scfm)

11.8 l/s (25 scfm) nominal

Air Supply Temperature

+20° to +25°C; +22°C nominal

Operating Environment

Operating Temperature

+20° to +28°C; +23°C nominal

Humidity

0 to 60%; 45% nominal

Service and Safety

Refrigerants

HCFC and CFC-free, non-toxic, non-flammable

Serviceability

Auto-diagnostics and field replaceable modules

Over Temperature Protection

+230°C (factory set): Operator can set high and low air temperature limits

Weight and Dimensions

Dimensions

Base

Width: 61.0 cm (24 in.), Depth: 72.4 cm (28.5 in.)

Height: 108 cm (42.5 in.)

System Weight

Not packed: 236 kg (520 lbs.)

Packed: 365 kg (805 lbs.)

Mobility

Four static dissipative, swivel caster wheels

Max. Reach

160.0cm (63 in.)

Max. Operating Height

130.3 cm (51.2 in.) - extended height option: 188.0 (74.0 in.)

Min. Operating Height

69.1 cm (27.2 in.) - extended height option: 81.3 (32.0 in.)

Noise Level

<65dBA

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E