Description
|
Compact ESD simulator with microprocessor based, large touch-sensitive LCD panel, built-in HV relay for contact-discharge, mains operation
|
Basics
|
Discharge pistol, high voltage base unit, mains power supply adapter, (100 to 250 VAC), discharge network 150 pF/330 Ω, air-and contact-discharge tips, grounding cable, user manual
|
Pulse data
|
Standard: Conforms to IEC/EN 61000-4-2 (2001) Special: Interchangeable networks for other standards
|
Pulse networks
|
Network 150 pF/330 Ω as per IEC/EN 61000-4-2 (included)
Optional ISO 10605 networks 330 pF/330Ω 150 pF/2 kΩ and 330 pF/2 kΩ
Range of RC-networks for other standards:
R = 0 Ω to 20 kΩ
C = 50 to 2000 pF
|
Discharge
|
Air-discharge: 200 V to 30 kV (in 100 V steps)
Contact-discharge: 200 V to 30 kV (in 100 V steps)
|
Discharge tips
|
Ball and point as per IEC and specials; exchangeable by threaded cap
|
|
Charging voltage measurement: kV, accuracy better than ±5% (stabilized)
|
Discharge detection
|
Air-discharge only: indicated by the kV symbol being displayed in inverse, also acoustically in the single‘s operating mode.
Threshold level on/off.
|
Holding time
|
> 5 s (charging voltage ±5%)
|
Charge resistor
|
50 MΩ
|
Triggering
|
Trigger button in hand-grip
|
Operation
|
Via touch panel and microprocessor
|
Discharge modes
|
Air-discharge/contact-discharge
|
Polarity
|
Positive, negative/automatic change
|
Operating modes
|
Single/repetitive, Pulse counter 0 to 9999
Pre-select counter 0 to 9999
Repetition:
- 0.5/1/5/10/20 or 25 Hz (air)
- 0.5/1/5/10 or 20 Hz (contact)
- Continuous operation
|
Auto-shut-off
|
After 15 minutes idle time (without loss of the test parameters)
|
Display
|
LCD panel showing: charging voltage, breakdown event, polarity, air-/contact-discharge, counter/preselect counter content
|
Weight
|
Discharge pistol (w/o cable): 1.2 kg (2.6 lbs) approx.
Base unit: 5.7 kg (12.6 lbs) approx.
|
Ambient conditions
|
Operation:
+5 to +40°C
20 to 80% r.h. (non-condensing)
68 to 106 kPa
|