The MD 300 probe has been specially designed to verify surge current pulses as specified in IEC/EN 61000-4-5, ANSI C62.41 and their derivates.
The main advantage of the MD 300 current probe is, that the measuring system is physically isolated from the circuit under test.
The MD 300 current probe is ready to use as coming along with pre-mounted coaxial cable. The BNC-end plug needs to be connected to the high-impedance input or 50 Ω input of an ordinary memory oscilloscope. Then the conductor carrying the surge current to be measured is passed through the hole in the current probe. The resulting voltage wave shape on the oscilloscope will then be an authenticable reproduction of the actual current wave shape within the given accuracy.
- Specially designed for surge current pulse measurements
- Probe is physically isolated from the circuit
- Ready to use
- No additional attenuators needed
Parameter
|
Value
|
Maximum peak current
|
5000 Amp
|
Maximum RMS current
|
>63 Amp
|
Nominal ratio
|
500:1 (into 1 MΩ system)
1000:1 (into 50 Ω system)
|
Sensitivity
|
0.002 V/Amp (1 MΩ system)
0.001 V/Amp (50 Ω system)
|
Standards
CISPR 11 |
CISPR 14 - 1 |
CISPR 14 - 2 |
CISPR 14 - 2 |
CISPR 16 - 1 |
CISPR 16 - 2 |
CISPR 16 - 3 |
CISPR 16 - 4 |
CISPR 22 |
CISPR 24 |
IEC/TR EN 61000 - 1 - 1 |
IEC/TR EN 61000 - 2 - 1 |
IEC/TR EN 61000 - 2 - 3 |
IEC EN 61000 - 3 - 2 |
IEC EN 61000 - 3 - 4 |
IEC/TC EN 61000 - 3 - 5 |
IEC EN 61000 - 4 - 2 |
IEC EN 61000 - 4 - 3 |
IEC EN 61000 - 4 - 4 |
IEC EN 61000 - 4 - 5 |
IEC EN 61000 - 4 - 6 |
IEC EN 61000 - 4 - 7 |
IEC EN 61000 - 4 - 8 |
IEC EN 61000 - 4 - 9 |
IEC EN 61000 - 4 - 11 |
EN 50 081 part 1 |
EN 50 081 part 2 |
EN 55 011 |
EN 55 013 |
EN 55 014 |
EN 55 015 |
EN 55 020 |
EN 55 022 |
EN 55 024 |
EN 50 082 part 1 |
EN 50 082 part 2 |
EN 50 093 |
FCC Part 15 |
MIL-STD - 461E |
|