Temperature Humidity Cycling Test Chambers
Tenney's temperature and humidity cycling test chambers are capable of simulating a wide range of temperature or temperature and humidity conditions. The ten cubic foot work space will readily accept equipment as large as a 19"relay rack.
In keeping with the needs of today's lab, Tenney designed these temperature humidity cycling test chambers to have a compact exterior that maximizes the interior workspace.
- Watlow F4 controller
- Solid state humidity sensor
- Platinum RTD temperature sensor
- Heavy duty leveling casters
- Nichrome wire air heater
- Solid state pressure switches
- Backup safety thermal cutoff
- Chamber vent port
- Communications - RS485 and RS232
Controller – Watlow F4
- Single or Dual Channel
- Profiling: Forty Profiles, 256 Steps
- Automatic & Manual Control
- 1 Input: Universal Type (RTD is Std.)
- 4 Digital Inputs
- 2 Auxiliary Analog Inputs - Optional
- 2 Control Outputs: Time Proportioned, On/Off, or 4-20 ma
- 2 Alarm Outputs
- 8 Digital Event Outputs
- 2 Retransmit Outputs - Optional
- Communications: RS232 and RS485
Model
|
TC10S
|
TC10C
|
TC10RS
|
TC10RC
|
Cubic Feet
|
10
|
Liters
|
283
|
Work Space (W x D x H)
|
24" x 27.5" x 28" / 61cm x 69cm x 71cm
|
Exterior (nominal) (W x D x H)
|
33.625" x 59.121" x 73.888" / 86cm x 151cm x 188cm
|
Unit Weight (LBS)
|
1000
|
1154
|
1000
|
1154
|
Temperature Range
|
-35°C
|
-73°C
|
-35°C
|
-73°C
|
|
200°C
|
200°C
|
200°C
|
200°C
|
Humidity Range*
|
-
|
-
|
20% - 95%RH
|
20% - 95%RH
|
Cooling Rate (empty chamber) Min (C/min)
|
180 to -65°C
|
-
|
77 (3.1)
|
-
|
77 (3.1)
|
180 to -35°C
|
71 (3)
|
56 (3.8)
|
71 (3)
|
56 (3.8)
|
125 to -35°C
|
60 (2.6)
|
41 (3.9)
|
60 (2.6)
|
41 (3.9)
|
85 to -35°C
|
53 (2.2)
|
30 (4)
|
53 (2.2)
|
30 (4)
|
85 to -20°C
|
28 (3.7)
|
25 (4.2)
|
28 (3.7)
|
25 (4.2)
|
ambient to -35°C
|
31 (1.9)
|
16 (3.7)
|
31 (1.9)
|
16 (3.7)
|
ambient to -68°C
|
-
|
38 (2.4)
|
-
|
38 (2.4)
|
Heating Rate (empty chamber) Min (C/min)
|
-65 to 180°C
|
-
|
35 (7)
|
-
|
35 (7)
|
-35 to 180°C
|
30 (7.1)
|
31 (6.9)
|
30 (7.1)
|
31 (6.9)
|
-35 to 125°C
|
20 (8)
|
20 (8)
|
20 (8)
|
20 (8)
|
-20 to 85°C
|
11 (9.5)
|
13 (8)
|
11 (9.5)
|
13 (8)
|
-35 to 85°C
|
10 (12)
|
14 (8.5)
|
10 (12)
|
14 (8.5)
|
-35°C to ambient
|
6 (10)
|
7 (8.5)
|
6 (10)
|
7 (8.5)
|
-68°C to ambient
|
-
|
10 (9.3)
|
-
|
10 (9.3)
|
Live Load Watts @ -18°C
|
950
|
1450
|
950
|
1450
|
Live Load Watts @ -40°C
|
-
|
1050
|
-
|
1050
|
Live Load Watts @ -54°C
|
-
|
800
|
-
|
800
|
Temperature Control
|
±0.3°C after stabilization
|
Temperature Uniformity
|
±3°C @ 65°C
|
Humidity Control
|
-
|
-
|
±2%RH
|
±2%RH
|
Air Flow Rate
|
275 CFM
|
Heat Capacity
|
3KW
|
Humidifier Capacity
|
-
|
-
|
2KW
|
2KW
|
Refrigeration
|
2HP
|
(2) 2HP
|
2HP
|
(2) 2HP
|
Condenser
|
Air Cooled - Std Water Cooled Optional
|
Power
|
208V - 230V 1P 60Hz
|
Amp - Fuse
|
40
|
60
|
50
|
60
|
Amp - Under Load
|
32
|
45
|
40
|
45
|
Power
|
460V 1P 60Hz
|
Amp - Fuse
|
20
|
30
|
30
|
30
|
Amp - Under Load
|
16
|
23
|
20
|
23
|
Performance estimates based on 230v 1P 60Hz
*Humidity Range: based on temperature range 20°C to 85°C limited by 3°C dewpoint
Standards
CISPR 11 |
CISPR 14 - 1 |
CISPR 14 - 2 |
CISPR 14 - 2 |
CISPR 16 - 1 |
CISPR 16 - 2 |
CISPR 16 - 3 |
CISPR 16 - 4 |
CISPR 22 |
CISPR 24 |
IEC/TR EN 61000 - 1 - 1 |
IEC/TR EN 61000 - 2 - 1 |
IEC/TR EN 61000 - 2 - 3 |
IEC EN 61000 - 3 - 2 |
IEC EN 61000 - 3 - 4 |
IEC/TC EN 61000 - 3 - 5 |
IEC EN 61000 - 4 - 2 |
IEC EN 61000 - 4 - 3 |
IEC EN 61000 - 4 - 4 |
IEC EN 61000 - 4 - 5 |
IEC EN 61000 - 4 - 6 |
IEC EN 61000 - 4 - 7 |
IEC EN 61000 - 4 - 8 |
IEC EN 61000 - 4 - 9 |
IEC EN 61000 - 4 - 11 |
EN 50 081 part 1 |
EN 50 081 part 2 |
EN 55 011 |
EN 55 013 |
EN 55 014 |
EN 55 015 |
EN 55 020 |
EN 55 022 |
EN 55 024 |
EN 50 082 part 1 |
EN 50 082 part 2 |
EN 50 093 |
FCC Part 15 |
MIL-STD - 461E |
|