QuadTech Guardian DC Hipot Tester 12 kV

Description

The Guardian 500VA, 10kVAC, 12kVDC and 20kVDC are four different Hipot Testers designed for specific electrical safety test applications. The Guardian 500VA is intended for high leakage current measurements to 100mA, the Guardian 10kVAC for high voltage AC testing and the Guardian 12kVDC and 20kVDC for high voltage DC testing. The Guardian 500VA, 12kVDC and 20kVDC instruments also measure insulation resistance (IR). Selecting the right instrument for the specific application provides a cost effective solution to special electrical safety compliance testing.

The Guardian 12kVDC Hipot Tester performs DC Hipot Testing over the voltage range from 100 to 12,000V DC with a maximum leakage current output of 10mA. Additionally, the 12kV tester provides Insulation Resistance measurement capability over a range of 100kΩ to 50GΩ at test voltages from 100 to 1000V DC. The new improved discharge function allows for fast controlled discharge of the DUT.

Features

  • DC Hipot Voltage 100 to 12,000V, Output Current to 10mA
  • IR Measurements 100kΩ to 50GΩ, Test Voltage 50 to 1000VDC
  • Programmable Ramp and Test Times
  • Fast Cutoff and Quick Discharge of Device Under Test
  • Programmable Hi/Lo Limits for Pass/Fail Testing
  • LCD Display for Menu Programming and Test Results
  • Internal Storage and Recall of 50 Test Setups, 10 Steps Each
  • Front and Rear Panel Output Connections
  • Front Panel Lockout Prevents Changes to the Test Setup
  • Automatic Offset to Zero Out Stray Leakage Current
  • Remote I/O Interface Included Standard
  • RS-232 Interface Optional
  • IEEE-488 and Printer Interfaces
  • Optional External 8 Channel Scanner Optional

Specifications

Guardian 12kVDC

DC Output Voltage

Range

100V to 12kV DC, 10V resolution

Regulation

<(1% +10V) at Rated Load

Voltage Display

Accuracy

 ±(1% of reading + 10V)

 

Resolution

10 Volt

DC Current Display

Range

0.01mA to 10mA DC

Resolution

0.0001mA

Accuracy

±(1% or reading + 5cnt)

High/Low Limit Test

0.1μA to 10mA DC

Low limit can be turned OFF

Insulation Resistance

Range

100kΩ - 50GΩ

Accuracy

±5% to ±15% depending upon V and R

Voltage Range

100V to 1000V DC

Voltage Accuracy

± (5% of setting + 5V)

High/Low Limit Test

100kΩ - 50GΩ

Low limit can be turned OFF

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E