»  Categories   »      Insulation Meter   »    Ground Bond Meter   »    Electrical Safety Tester   »    DC Hipot   »    AC Hipot   »    Environmental   »    Analyzer  

image enlarge

QuadTech Guardian 6000 Plus Electrical Safety Analyzer

Description

The Guardian 6000 Plus is able to perform several dielectric tests in one unit including AC/DC Hipot, Insulation Resistance, Ground Bond, Open/Short Check and optional Leakage Current for compliance testing of IEC, UL, TUV, CSA, EN and other standards including IEC60601-1, UL2601-1, IEC950, UL544, UL60950. The time-saving TwinPort feature performs Hipot and Ground Bond simultaneously cutting time for these tests in half. The open / short check function (OSC) checks for an open connection or a DUT short during testing. Open connections may cause a bad DUT to test OK. Screening for shorts prevents damage to the equipment saving potential costs that may occur. These patented features are only available in safety products manufactured by Chroma.

Features

  • Five instruments in one: AC Hipot , DC Hipot, Insulation Resistance, Ground Bond and Leakage Current (Option)
  • Twin-Port™ feature (Patented)
  • Programmable output voltage to 5kV AC and 6kV DC
  • Insulation resistance to 50GO/1000V DC
  • Ground bond up to 30A (Option up to 40A / 60A)
  • Open/Short check (OSC) function
  • ARC detection (Flashover)
  • Password Protected front panel lockout
  • Storage of 50 Tests Setups with 100 Steps per setup
  • Optional dynamic leakage current auto scanning
  • Standard RS-232 Interface
  • Standard GB OFF.S.ET KIT, SCANNER Interface
  • Optional GPIB Interface

Specifications

Withstanding Voltage Test

Output Voltage

DC:0.05 ~ 6kV , AC : 0.05 ~ 5kV

Load Regulation

±(1% of reading +0.1% of range)

Voltage Regulation

2V

Voltage Accuracy

±(1% of reading +0.1% of range)

Cutoff Current

DC : 20mA , AC : 40mA

Current Resolution

0.1 μA DC

Current Accuracy

±(1% of reading +0.1% of range)

Output Frequency

50Hz / 60Hz

Test Time

0.3 ~ 999 sec , continue

Ramp Time

0.1 ~ 99.9 sec, Off

Fall Time

0.1 ~ 99.9 sec, Off

Waveform

Sine wave

Insulation Resistance Test

Output Voltage

DC : 0.05 ~ 1kV

Voltage Resolution

2V

Voltage Accuracy

±(2% of reading +0.5% of range)

IR Range

1MΩ ~ 50GΩ

Resistance Resolution

0.1MΩ

Resistance Accuracy

5% typical

Ground Bond Test

Output Current

AC : 1 ~ 30A (40A, 60A optional)

Current Accuracy

±(1% of reading +0.2% of range)

GR Range

10mΩ ~ 510mΩ

Resistance Resolution

0.1mΩ

Resistance Accuracy

± (1% of reading + 0.1% of full scale)

Test Method

4 wires

General

Operation Environment Temperature

0°C ~ 40°C, Humidity : ± 80 % RH

Power Consumption

No load : < 100 W With rated load : 800 W

Power Requirements

90~132Vac or 198~264Vac, 47~66Hz

Weight Approx.

20KG(19032) 25KG (19032-P)

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E