IEC / EN and ANSI standards cover burst testing of single and three phase AC and DC power ports. They include recommendations for the test equipment, the test set-ups and the test procedures. The IEC 61000-4-4 Edition 2: 2004 now defines the impulse shape not only at the generator output but also at the CDN output.
This Edition 2 requires a pure common mode coupling only. This means that the burst signals are coupled simultaneously into all paths at the same time. Because of this no coupling path selection during testing is necessary. This speeds up the test time.
The FP-EFT 32M couples the burst into a EUT while preventing the impulses from polluting the three phase power supply.
The FP-EFT 32M has manual coupling path selection to verify the functionality of each path as recommended in the IEC 61000-4-4 Edition 2: 2004. This can also be used to select different coupling paths according to ANSI standards which require a path by path burst test.
Used together with the PEFT 4010, PEFT Junior or PEFT.1 generators, the FP-EFT 32M can be used to inject EFT/burst impulses on the three phases supply.
- 690Vac phase-phase, 400Vac phase-neutral
- 32A per phase EUT current
- Maximum EFT voltage 8kV
- manual coupling path selection
- IEC/EN 61000-4-4 edition 1&2
- ANSI C62.41 & C37.90.1
- Phase angle synchronization possible
Applications
- Three phase power line systems
- IEC/ EN 61000-4-4 Power lines
- ANSI C62.41 & C.37.90.1 Power lines
- Many IEC & EN Product standards
Maximum Impulse Voltage
|
8kV Burst
|
Maximum AC Voltage
|
690V (phase – phase)
400V (phase – neutral)
|
Maximum DC Voltage
|
110V
|
Maximum AC Current
|
32A
|
Maximum DC Current
|
32A
|
Mains Connections
|
4mm safety banana sockets
|
EUT connections
|
4mm safety banana sockets
|
Phase Sync. source
|
fixed between L3 and PE
|
Coupling Paths
|
Manual Selection
|
Residual pulse voltage at line input
|
Max. 10% of the applied impulse amplitude
|
Dimensions (W x H x D)
|
45 x 20 x 50 cm
|
Weight
|
10 kg
|
Standards
CISPR 11 |
CISPR 14 - 1 |
CISPR 14 - 2 |
CISPR 14 - 2 |
CISPR 16 - 1 |
CISPR 16 - 2 |
CISPR 16 - 3 |
CISPR 16 - 4 |
CISPR 22 |
CISPR 24 |
IEC/TR EN 61000 - 1 - 1 |
IEC/TR EN 61000 - 2 - 1 |
IEC/TR EN 61000 - 2 - 3 |
IEC EN 61000 - 3 - 2 |
IEC EN 61000 - 3 - 4 |
IEC/TC EN 61000 - 3 - 5 |
IEC EN 61000 - 4 - 2 |
IEC EN 61000 - 4 - 3 |
IEC EN 61000 - 4 - 4 |
IEC EN 61000 - 4 - 5 |
IEC EN 61000 - 4 - 6 |
IEC EN 61000 - 4 - 7 |
IEC EN 61000 - 4 - 8 |
IEC EN 61000 - 4 - 9 |
IEC EN 61000 - 4 - 11 |
EN 50 081 part 1 |
EN 50 081 part 2 |
EN 55 011 |
EN 55 013 |
EN 55 014 |
EN 55 015 |
EN 55 020 |
EN 55 022 |
EN 55 024 |
EN 50 082 part 1 |
EN 50 082 part 2 |
EN 50 093 |
FCC Part 15 |
MIL-STD - 461E |
|