GENERAL DATA
|
Control power
|
85 V - 264 V
50/60 Hz
|
User test storage
|
unlimited
|
Remote interface
|
Ethernet, RJ45
|
Display
|
7“/ 800x480 / 24 bit with touch-screen
|
USB
|
for USB memory stick
|
AUX. interface
|
D-sub 37p for external CDN, external transformer etc.
|
External trigger input
|
5 V TTL
|
Synch input
|
BNC, 10 V – 264 V AC
|
Trigger output
|
5 V TTL
|
External start / stop input
|
5 V TTL, starts / stops predefined test sequence
|
EUT failed input
|
5 V TTL
|
Analog output
|
0 – 10 V, for use with external options
|
Warning lamp output
|
2 x 24 V / 1 A DC
|
Safety circuit
|
stops the test when unlocked
|
Dimensions
|
19“/ 4U
|
(W x H x D)
|
(45 x 18 x 49 cm)
|
Weight
|
30 kg
|
IEC / EN 61000-4-4 EDITION 2 EFT / BURST
|
Output voltage
|
0.2 – 5.0 kV ±10% at coaxial output
|
Spike frequency
|
1 Hz – 1 MHz
|
Polarity
|
pos / neg / alternate
|
Burst duration
|
10 us – 1 s
|
Output impedance
|
50 Ohms
|
Burst period
|
1 ms – 10 s
|
Rise time
|
5 ns ±30%
|
Test time
|
1 s– 999 minutes
|
Impulse duration
|
50 ns ±30% at 50 Ohm
50 ns –15 +100 ns at 1000 Ohm
|
Trigger
|
automatic, manual,
external trigger input
|
Burst mode
|
normal, continuous, real, random
|
Integrated single phase coupling / decoupling network
|
264 V AC / 16 A
220 V DC / 10 A
|
IEC / EN 61000-4-5 EDITION 2 SURGE COMBINATION WAVE
|
Output voltage
|
0.2 – 5.0 kV ±10% Output current 0.1 – 2.5 kA ±10%
|
Voltage rise time
|
1.2 us ±30% Current rise time 8 us ±20%
|
Voltage duration
|
50 us ±20% Current duration 20 us ±20%
|
Polarity
|
pos / neg / alternate
|
Integrated single phase CDN
|
264 V AC / 16 A
220 V DC / 10 A
|
Output impedance
|
2 Ohms
|
Phase synch
|
0 – 359° with 1° steps or asynchronous mode
|
Impulse trigger
|
automatic 5 s – 99999 s
manual external trigger input
|
Counter preselect counter
|
1 – 10000 / infinite 100000
|
Peak voltage monitor
|
BNC output: 1000:1
display: 3 digits
|
Peak current monitor
|
BNC output: 1 kA/V
display: 3 digits
|
IEC / EN 61000-4-11 EDITION 2 AND IEC / EN 61000-4-29 DIPS & INTERRUPTS
|
Max. voltage
|
264 V AC/DC Interrupt time 0.1 period – 9999 periods
0.02 ms – 999 minutes
|
Max. current
|
16 A AC/DC continuous
20 A for 5 s
40 A for 3 s
500 A inrush current
|
Interval time
|
1 period – 9999 periods synch
0.04 ms – 999 minutes asynch
|
Trigger
|
automatic
Manual external trigger input
|
Test time
|
1 s – 999 minutes
infinite
|
Interrupt dip level
|
0%
0% – 99% with external voltage source
|
Phase synch
|
0 – 359° 16 / 40 / 50 / 60 Hz
asynchronous mode
|
RMS voltage monitor
|
BNC output: 100:1
display: 4 digits
|
RMS current monitor
|
BNC output: 10 A/V
display: 4 digits
|