ETS-Lindgren MF-130D RF Test Set

Description

The RF Attenuation Meter MF-130D was designed to simplify the checking of the performance of a shielded enclosure.

With the possibility of generating eight frequencies, the engineer is able to check the shielding effectiveness very quickly. The measurement can be done at four frequencies: 8 MHz, 16 MHz, 32 MHz and 64 MHz. The measurements can be performed according to the following standards: MIL-STD-285 NSA 65-6 IEEE STD-299 MF-130D consists of portable, NiCd battery operated transmitter and receiver. Frequency and attenuation calibration is automatically operated by a microprocessor. The receiver is equipped with an LCD-display which shows the attenuation in dB at the normal operational mode. An alarm sound is heard when the attenuation is less than the selected level at the preset limit mode.

Saving and downloading the measured data You can save 63 attenuation levels at each frequency. You can download these saved values to a computer through a serial (RS-232) connection. The MF-130D Attenuation Meter set is provided with two diskettes. These diskettes are for downloading the measuring information into a computer. One diskette is for a Windows 3.1x operating system and the other is for a Windows 95 operating system.

Measuring magnetic attenuation according to NSA 65-6 / IEEE STD-299: Place the receiver and the transmitter as shown in the figure above at the point to be measured with the receiver inside the enclosure and the transmitter outside. The attenuation value may now be read from the meter. Measuring magnetic attenuation according to MIL-STD-285: Place the receiver and the transmitter as shown in the figure above. Turn the antenna of the receiver and the transmitter to find out the minimum attenuation reading.

Features

  • Software to download test data
  • One transmit and one receive unit
  • Two battery chargers
  • Carrying Case

Specifications

Electrical Specifications

MF-130D

LF-antenna

HS-antenna

This device is in compliance

with the following standards

EN 50081-1 (1992)

EN 50082-1 (1992)

EN 50081-1 (1992)

EN 50082-1 (1992)

EN 50081-1 (1992)

EN 50082-1 (1992)

Frequency ranges

N / A

10 kHz

156 kHz

1 MHz

10 MHz

8 MHz

16 MHz

32 MHz

64 MHz

Frequency accuracy

max. (5 ppm / 10 Hz)

N / A

N / A

Dynamic range    MIL-STD-285

NSA 65-6

IEEE STD-299

EN 50147-1

130 dB

130 dB

130 dB

130 dB

120 dB

130 dB

130 dB

130 dB

110 dB

120 dB

120 dB

120 dB

Attenuation accuracy ±

2 dB

N / A

N / A

Output power

N / A

2 W max.

2 W max.

Antenna

N / A

1 layer PCB antenna

1 layer PCB antenna

Display

128 x 64 LCD matrix display with background light

Memory

63 memory locations for attenuation value for each frequency

Preset functions

Attenuation limit in dB

Calibration

Auto calibration

Power supply, both

6 x 1.2 Volt NiCd battery (D)

Battery life

Transmitter 5 to 7 hours

Receiver 10 to 12 hours

Physical specifications

Case dimensions

530 mm x 190 mm x 400 mm (L x W x H)

Temperature range

+5° C to +40° C (41° F to 103° F)

Storage temperature

-5° C to +45° C (23° F to 113° F)

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E