Input
|
Voltage Ranges
|
Factory configured 187 to 264 Vrms, 3ø L-L (3 wire), or 342 to 457 Vrms, 3ø L-L (4 wire). A chassis ground is also required.
|
Power Factor
|
PFC Input, 0.99PF
|
Frequency Range
|
47 to 440 Hz
|
Efficiency
|
70% min, at full load
|
Ride Through
|
70% min, at full load
|
Output
|
Power
|
5550 VA: 1ø or 3ø (systems up to 33,000 VA)
|
AC or DC Output Voltage
|
0 to156 Vrms L-N, low range; 0 to 312 Vrms L-N, high range
|
Current Per Phase
|
13/16A to 135V in 156V range; 6.5/13A to 270V in 312V range per 1750/1850 VA module. Note: Higher currents may be achieved in 10.5-21 kVA systems.
|
Power Factor of Load
|
0 lagging to 0 leading (0-unity)
|
Crest Factor
|
3.25:1 (peak output current to rms output current)
|
Frequency Range
|
Specifications apply DC, 40Hz to 5kHz. For output frequencies greater than 1 kHz, the max slew rate allowed is 1 kHz per second.
|
Max Total Harmonic Distortion
|
(Full Linear Load or No Load): 0.25% max, 40 to 100 Hz; 0.5% max to 500 Hz; and 1% max to 1 kHz plus 1%/kHz to 5 kHz
|
AC Noise Level
|
>60 dB rms below full output voltage
|
Amplitude Stability With Remote Sense
|
±0.1% of full scale over 24 hours at constant line, load and temperature
|
Line Regulation
|
(DC, or 40 Hz to 5 kHz): ±0.015% of full scale for a ±10% input line change
|
Load Regulation
|
±0.025% of full scale voltage for a full resistive load to no load; above 1 kHz, add ±0%/kHz
|
Voltage Accuracy
|
±0.1% of range. Above 1 kHz, add 0.2%/kHz. Add ±0.1% of full scale for "AC PLUS DC" mode. Valid for 5 to 156 Vrms and 10 to 312 Vrms at 25°C
|
Voltage Resolution
|
1mV (0.1V) Full Scale
|
Frequency Accuracy
|
±0.01% at 25°C ±0.001%/°C
|
Frequency Resolution
|
40 Hz to 81.91 Hz (0.01 Hz) 81.90 Hz to 819 Hz (0.01 Hz) 820 Hz to 5000 Hz (1 Hz)
|
Phase Accuracy, Phase-to-Phase Balanced Linear Resistive Load
|
±1% of Programmed value
|
Phase Angle Resolution
|
0.1°
|
Remote Output Voltage Sense
|
5 Vrms total lead drop, max
|
Common
|
Standard Features
|
• 1ø to 3ø programmable • BNC outputs for scope viewing of waveforms (1 MΩLoad Drive)
|
Interface
|
IEEE-488.2 interface, RS232, USB, Optional LAN
|
Protocol
|
SCPI protocol
|
WaveForm trigger output
|
(1 MΩLoad Drive)
|
Sync OUT
|
User programmed for: Cycle start, all cycles. Segment start, all or selected segments. For loads ≥2 kΩ: V out ≤1V low state; V out ≥2.4V high state
|
Clock/lock
|
Clock - pulses at programmed frequency. For loads ≥2 kΩV out: ≤1V low state; V out ≥2.4V high state Lock - locks output to input 'TTL' frequency; signal needs to supply pull down current of 15 mA with voltage drop of ≤0.6V; no pull up needed
|
External Amplitude Modulation
|
0 to 5 Vrms provides 0 to ≥20% output amplitude modulation
|
External Drive
|
Normal amplifier, 0 to 5 Vrms (DC to 5 kHz) or ±5 VDC input for zero to full voltage output
|
External Gain Control
|
0 to ±7.07 VDC provides zero to full output
|
External Input Impedance
|
≥30 kΩ
|
Environmental
|
Operating Temperature
|
0°C to 45°C (32°F to 113°F)
|
Storage Temperature
|
-40°C to 70°C (-40°F to 158°F)
|
Cooling
|
Air is drawn in from the top, bottom, and sides and exhausted through the rear
|
Humidity (Non-condensing)
|
0 to 85% at 25°C (77°F); derate to 50% at 40°C (104°F)
|
Altitude
|
Operating 10,000 ft, non operating 40,000 ft
|
Physical
|
Dimensions
|
Width: 19" (483 mm) Height: 8.75" (222 mm) Depth: 23.5" (597 mm)
|
Weight
|
CSW 5550 - 127 lbs (57.5 kg)
|
Shipping Weight - US
|
CSW - 200 lbs (91 kg)
|
Note: Multi Chassis systems, dimension and weight are approximately x2, x3 and x4 CSW Specifications
|
Measurements - Standard (AC Measurements)
|
Parameter
|
Frequency
|
Phase
|
Voltage (AC rms)
|
Current (AC rms)
|
Range
|
40-81.91 Hz 82.0-819.1 Hz > 819 Hz
|
40-100 Hz 100-1000 Hz
|
0-300 V
|
0-50 A
|
Accuracy* (±) 1 ø mode (-1)
|
0.1% + 1 digit
|
0.5°
|
0.5% + 250 mV
|
0.1% + 150 mA
|
Accuracy* (±) 3 ø mode (-3)
|
.01 Hz / 0.1 Hz / 1 Hz
|
2°
|
0.5% + 250 mV
|
0.1% + 50 mA
|
Resolution*
|
.01 Hz / 0.1 Hz / 1 Hz
|
0.1° / 1°
|
10 mV
|
1 mA
|
Parameter
|
Real Power
|
Apparent Power
|
Power Factor
|
Range
|
0-6 kW
|
0-6 kVA
|
0.00-1.00
|
Accuracy* (±) 1 ø mode (-1)
|
0.15% + 9 W
|
0.15% + 9 VA
|
0.03
|
Accuracy* (±) 3 ø mode (-3)
|
0.15% + 3 W
|
0.15% + 3 VA
|
0.01
|
Resolution*
|
1 W
|
1 VA
|
0.01
|
* Accuracy specifications are in % of reading and apply above 100 counts. For multi-chassis configurations, current, power range and accuracy specifications are times three. Power factor accuracy applies for PF > 0.5 and VA > 50% of max. Frequency measurement specification valid for output > 30 Vrms.
|
Harmonic Measurements
|
Parameter
|
Range
|
Accuracy ( ± )
|
Resolution
|
Frequency fundamental
|
16.00 - 1000 Hz
|
2 counts
|
0.01 Hz to 1 Hz
|
Frequency harmonics
|
32.00 Hz - 16 kHz
|
2º typ.
|
0.5º
|
Voltage
|
Fundamental
|
0.25V
|
0.01V
|
Harmonic 2 - 50
|
0.25V + 0.1% + 0.1%/kHz
|
0.01V
|
Current
|
Fundamental
|
0.05A
|
0.01A
|
Harmonic 2 - 50
|
0.05A + 0.1% + 0.1%/kHz
|
0.01A
|
Harmonics frequency range in three-phase mode is 32 Hz - 16 kHz. Accuracy specifications are multiplied by the number of power sources in multi-source systems with the 3-phase (3Ø) or the number of sources times 3 in the 1-phase (1Ø) mode. Measurement bandwidth is limited to 16 Khz.
|
Model
|
CSW5550
|
CSW11100
|
CSW16650
|
CSW22200
|
Multiplier 3Ø/1 Ø
|
1/3
|
2/6
|
3/9
|
4/12
|
Model
|
CSW17750
|
CSW33300
|
CSW38850
|
CSW44400
|
Multiplier 3Ø/1 Ø
|
5/15
|
6/18
|
7/21
|
8/24
|
Protection And Safety
|
Overvoltage Shutdown
|
Programmable for 15V to 255V peak, 156V range; 30V to 510V peak, 312V range
|
Programmable Current Limit Shutdown
|
Settable to 0.01 ARMS Resolution
|
Programmable Current Limit with Timed Shutdown
|
Settable to 1% of range: the timeout is settable from 100 ms to 10s.
|
Over temperature Shutdown
|
Automatic, not programmable
|
Regulatory Compliance
|
• EN 61010 • EN 55011 • UL 3111 • EN 50082-2 • EN 61000-4-3, EN 61000-4-4 • FCC Part 15, Class A • CE Mark
|