Input Specifications
|
Voltage
|
115/230 VAC ± 10%, user selection
|
Frequency
|
50/60 Hz ± 5%
|
Fuse
|
2 A Slow Blow 250 VAC
|
Line Conditions
|
Reverse Power
|
Switch Reverse polarity switch for normal condition
|
Neutral Switch
|
Neutral switch on/off selection for single fault
|
Ground Switch
|
Ground switch on/off selection for class I single fault
|
Probe Settings
|
Surface to Surface
|
(PH - PL)
|
Surface to Line
|
(PH - L)
|
Ground to Line
|
(G - L)
|
Leakage Limit Settings
|
Touch Current
High/Low
Limit (RMS)
|
Range
|
0.0 μA - 999.9 μA / 1000 μA - 9999 μA /10.00 mA - 20.00 mA
|
Resolution
|
0.1 μA/1 μA/0.01 mA
|
Touch Current High/Low
Limit (PEAK)
|
Range
|
0.0 μA - 999.9 μA/ 1000 μA - 9999 μA / 10.00 mA - 30.00 mA
|
Resolution
|
0.1 μA/1 μA/0.01 mA
|
Display
|
Touch Current
Display (RMS)
|
Range
|
0.0 μA - 550 μA, frequency DC, 15 Hz - 1 MHz
|
Resolution
|
0.1 μA
|
Accuracy
|
DC: 15 Hz < f < 100 kHz: ±(2% of reading + 3 counts) 100 kHz < f < 1 MHz: ±5% of reading, (10.0 μA - 999.9 μA)
|
Range
|
400 μA - 8500 μA, frequency DC, 15 Hz - 1 MHz
|
Resolution
|
1 μA
|
Accuracy
|
Accuracy: DC: 15 Hz < f < 100 kHz: ±(2% of reading + 3 counts) 100 kHz < f < 1 MHz: ±5% of reading, (10 μA - 8500 μA)
|
Range
|
8.00 mA - 20.00 mA, frequency DC, 15 Hz - 1 MHz
|
Resolution
|
0.01 mA
|
Accuracy
|
DC: 15 Hz < f < 100 MHz: ±5% of reading, (0.01 mA -20.00 mA)
|
Touch Current Display (PEAK)
|
Range
|
0.0 μA - 550 μA, frequency DC - 1 MHz
|
Resolution
|
0.1 μA
|
Accuracy
|
DC : ±(2% of reading +3 counts) 15 Hz < f < 1 MHZ : ±10% of reading +2 μA
|
Range
|
400 μA - 8500 μA, frequency DC - 1 MHz
|
Resolution
|
1 μA
|
Accuracy
|
DC: ±(2% of reading +3 counts) 15 Hz < f < 1 MHz ±10% of reading +2 μA
|
Range
|
8.00 mA - 30.00 mA, frequency DC - 100 kHz
|
Resolution
|
0.01 mA
|
Accuracy
|
DC: ±(2% of reading +3 counts) 15 Hz < f < 100 kHz ±10% of reading +2 counts
|
General Specifications
|
Dimension (W x H x D)
|
16.93 x 5.24 x 11.81 in (430 x 133 x 300 mm)
|
Weight
|
26.45 lbs (12 kg)
|
Display
|
320 X 240 graphic LCD
|
Mechanical
|
Bench or rack mount with tilt up feet
|
Memory
|
50 Memories, 30 steps per each memory
File locations can link 900 steps max
|
Interface
|
USB/RS232 Standard, Ethernet, GPIB, Data Storage
(RS-485) Optional
|