»  Categories   »      CDN   »    EMC / EMI   »    Coupling / Decoupling (CDN)  

image enlarge

EM TEST CNV 508N4 Coupling/Decoupling Networks 4 Pairs / 8 Lines

Description

The CNV 508N4 series are special coupling/decoupling networks being used to perform "Electrical disturbance tests for measuring relays and protection equipment" in accordance to IEC 60255-22-1. A capacitance of 0.5uF is specified to couple the pulses on to the lines.

Each line has a decoupling inductor of >1.5mH to protect the auxiliary equipment from being subjected to the test pulse. The disturbance signal used for this test is a damped oscillatory pulse with a frequency of 1MHz, generated by the EM TEST OCS 500N6.

Depending on the configuration of the DUT several different tests set-ups are used to perform tests in differential mode (line-to-line of the same pair) or common mode (each pair to ground or one pair to all other pairs with these being grounded). The CNV 508N4 is automatically set by the OCS 500N6 selecting the desired coupling mode and lines to be coupled. The coupler offers capability to test up to 4 pairs/8 lines.

The test of measuring relays and protection equipment as outlined in IEC 60255-22-1 includes different coupling modes to test the DUT under different conditions. Each pair of lines is tested individually in differential mode. Common mode test is applied to each pair to ground and to each pair with the other pairs connected to ground.

Features

  • Selection of all coupling modes as per IEC 60255-22-1
  • Remote controlled by the OCS 500N6
  • Coupling capacitor 0.5uF
  • Decoupling inductor >1.5mH per line
  • Nominal line voltage max. 250V AC/DC
  • Nominal line current 4A AC/DC or 16A AC/DC

Specifications

OVERVIEW

CNV 508N4 MODELS

CNV 508N4

Coupling/decoupling network with 4A AC/DC per line

CNV 508N4.1

Coupling/decoupling network with 16A AC/DC per line

TECHNICAL DETAILS

CNV 508N4

Nominal voltage

250V AC/DC per line

Nominal current

4A AC/DC per line

Max. test voltage

3kV peak for 1MHz damped oscillatory wave

Source impedance 200 Ohm

CNV 508N4.1

Nominal voltage

250V AC/DC per line

Nominal current

16A AC/DC per line

Max. test voltage

3kV peak for 1MHz damped oscillatory wave

Source impedance

200 Ohm

COMMON DATA (ALL MODELS)

OUTPUTS

DUT connectors

4mm safety lab connectors

Indications

LEDs to indicate selected coupling mode and lines to be tested

COUPLING MODES AS PER IEC 60255-22-1

Differential Mode

Line-to-line within one pair

Common Mode #1

Test between an individual port and all others (ports not coupled are floating)

Common Mode #2

Test between an individual port and all others (ports not coupled are connected to ground via capacitor)

GENERAL DATA

Dimensions

19" 3HU

Weight

approx. 10kg  (CNV 508N4)

approx. 25kg (CNV 508N4.1)

Supply voltage

115/230V +10/-15%, 50/60Hz

Fuses

2 x T2AT (230V) or 2 x T4AT (115V)

CN interface

CN interface, 15pin SubD, for remote control by the OCS 500N6

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E