»  Categories   »      Stability Test Chamber   »    Environmental Chamber   »    Environmental   »    Chamber / Oven  

image enlarge

Blue M HRP Series High Temperature & High Humidity Test Chamber

Description

Continuous High Temperature and High Humidity Test Chamber

Blue M HRP Series Chambers provide high-range steady state humidity. These chambers are designed to deliver large volume high-temperature/high-humidity conditions. HRP chambers incorporate horizontal airflow and a one-pass airflow system for an easy-to-use, highly effective, temperature/humidity system. This air system uses room ambient air for cooling, dehumidification and chamber exhaust. The system controls the input of room temperature air, the mixing of room air and chamber air, and the exhaust volume.

Processing silicone rubber compounds and other products requires high-moisture content for thorough, rapid curing. These chambers are also used to test rigid and flexible membranes and for aggressive high-temperature, high-humidity tests such as 85°C (185°F)/85% humidity.

Features

  • Vapor pressure system
  • Horizontal air circulation
  • Separate temperature and humidity conditioning plenums
  • Solid state water levelers
  • Adjustable airflow
  • 100% recirculation to single pass system
  • Open air electric heaters
  • Vaisala Model HMM100 humidity module

Performance

  • Temperature: + 15°C above ambient to + 150°C (+ 302°F), +/- 1° after stabilization, measured at air inlet to workspace
  • Humidity: 10% to 95% RH in the dry bulb range from + 40°C to + 93°C (+ 104°F to + 199.4°F), measured at air inlet to workspace

Controller – Watlow F4

The Watlow F4 controller provides 24-step programming or may be used as a non-ramping manual controller.

  • Single Channel
  • Profiling: Forty Profiles, 256 Steps
  • Automatic & Manual Control
  • 1 Input: Universal Type (RTD is Std.)
  • 4 Digital Inputs
  • 2 Auxiliary Analog Inputs - Optional
  • 2 Control Outputs: Time Proportioned, On/Off, or 4-20 ma
  • 2 Alarm Outputs
  • 8 Digital Event Outputs
  • 2 Retransmit Outputs - Optional
  • Communications: EIA-232 and EIA-485 with Modbus RTU protocol

Specifications

Model #
-15 to +93°C

Inside Dimensions Inches (CM)

Overall Dimensions Inches (CM)

Full Load Amps

Voltage 60Hz AC

Cubic Feet Capacity (Liters)

W

D

H

W

D

H

HRP-09B

25 (64)

25 (64)

25 (64)

63 (160)

37 (94)

71 (180)

28

208V/1PH

9 (262)

HRP-09C

25 (64)

25 (64)

25 (64)

63 (160)

37 (94)

71 (180)

31

240V/1PH

9 (262)

HRP-13B

37 (94)

25 (64)

25 (64)

75 (191)

37 (94)

71 (180)

32

208V/1PH

13 (385)

HRP-13C

37 (94)

25 (64)

25 (64)

75 (191)

37 (94)

71 (180)

36

240V/1PH

13 (385)

HRP-27E

36 (91)

36 (91)

36 (91)

75 (191)

48 (122)

83 (211)

34

208V/3PH

27 (734)

HRP-27F

36 (91)

36 (91)

36 (91)

75 (191)

48 (122)

83 (211)

37

240V/3PH

27 (734)

HRP-27G

36 (91)

36 (91)

36 (91)

75 (191)

48 (122)

83 (211)

19

208V/3PH

27 (734)

HRP-64E

48 (122)

48 (122)

48 (122)

89 (226)

60 (152)

98 (249)

50

208V/3PH

64 (1816)

HRP-64F

48 (122)

48 (122)

48 (122)

89 (226)

60 (152)

98 (249)

56

240V/3PH

64 (1816)

HRP-64G

48 (122)

48 (122)

48 (122)

89 (226)

60 (152)

98 (249)

28

208V/3PH

64 (1816)

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E