Common
|
Meter Accuracy
|
1% of full scale + 1 count
|
Max. Voltage Differential from Output to Safety Ground
|
150 VDC
|
Remote Start/Stop and Interlock
|
TTL compatible input or 12-250 VAC (12-130 VDC) or a contact closure
|
Cooling
|
Internal fan, over temperature shutdown if internal heat sink exceeds set temperature
|
Remote Sense
|
The maximum allowed sense line drop is 4V per line (2V on the DCS 8/10V 1 kW/1.2 kW models and 1V/line for all 3 kW models). Line drop subtracts from the maximum available output voltage at full rated power.
|
Remote Programming
|
Enabled via external jumper on rear panel connector J3
|
Overvoltage Protection
|
Crowbar type adjustable from 5-110% of rated output using front panel control (local or remote program selectable via J3 jumper)
|
Remote Analog Programming Linearity
|
±1% Accuracy: ±5%
|
Regulatory
|
Certified to UL/CSA 61010 and IEC/EN 61010-1 CE Compliant (LVD and EMC directive)
|
Input: Voltage and Frequency
|
Output Power
|
200 - 250 VAC
|
100 - 132 VAC
|
1kW
|
single phase, 8A typical, 47- 63 Hz
|
single phase, 15A typical, 47-63 Hz, internal jumper selectable
|
Output: Voltage and Current
|
1kW Model
|
Voltage
|
Current
|
DCS 8-125E
|
0 - 8
|
0 - 125
|
DCS 10-100E
|
0 - 10
|
0 - 100
|
DCS 20-50E
|
0 - 20
|
0 - 50
|
DCS 33-33E
|
0 - 33
|
0 - 33
|
DCS 40-25E
|
0 - 40
|
0 - 25
|
DCS 50-20E
|
0 - 50
|
0 - 20
|
DCS 60-18E
|
0 - 60
|
0 - 18
|
DCS 80-13E
|
0 - 80
|
0 - 13
|
DCS 100-10E
|
0 - 100
|
0 - 10
|
DCS 150-7E
|
0 - 150
|
0 -7
|
DCS 300-3.5E
|
0 - 300
|
0 - 3.5
|
DCS 600-1.7E
|
0 - 600
|
0 - 1.7
|
Output
|
Voltage Resolution
|
0.02%
|
Ripple (mV rms or p-p)
|
See table
|
Regulation Line
|
For input voltage variation over the AC input voltage range, with constant rated load.
|
Load
|
For 0-100% load variation, with constant nominal line voltage.
|
Voltage
|
0.1%
|
Current
|
0.1%
|
Transient Response
|
Typically recovers in 500 μs (1 & 1.2 kW) or 1ms (3k W) to 1% of steady-state output voltage (within 1% of Vmax) for 70-100% or 100-70% load change.
|
Stability
|
±0.05% of maximum voltage or current over 8 hours after 30 minute warm-up time at fixed line, load and temperature
|
Efficiency
|
See table below Temperature Coefficient: 0.02%/°C of maximum output voltage; 0.03%/°C of maximum output current. Change in output per °C change in ambient temperature, with constant line and load
|
Physical
|
Dimensions
|
Width: 19” (482 mm) W, Depth 17.52” (445 mm)
Height: 1 kW and 1.2 kW: 1U or 1.72”
|
Weight & Shipping Weight
|
Weight
|
1 kW and 1.2 kW: 19 lbs. ( 8.6 kg)
|
Shipping Weight
|
1 kW and 1.2 kW: 24 lbs. (10.9 kg)
|
Environmental
|
Operating Temperature
|
0°C to 50°C (no derating)
|
Storage Temperature
|
-55°C to 85°C
|
Humidity Range
|
0 to 80% RH, non condensing
|
Programming Specifications
|
DCS Series 1 kW
|
Programming Accuracy M130/M131/M9C/M85 Options
|
Voltage 0.1% +
|
Current 0.1% +
|
OVP 0.5% +
|
DCS 8-125E
|
8mV
|
500mA
|
44mV
|
DCS 10-100E
|
10mV
|
400mA
|
55mV
|
DCS 20-50E
|
20mV
|
200mA
|
110mV
|
DCS 33-33E
|
33mV
|
132mA
|
182mV
|
DCS 40-25E
|
40mV
|
100mA
|
220mV
|
DCS 50-20E
|
50mV
|
80mA
|
275mV
|
DCS 60-18E
|
60mV
|
72mA
|
330mV
|
DCS 80-13E
|
80mV
|
52mA
|
440mV
|
DCS 100-10E
|
100mV
|
40mA
|
550mV
|
DCS 150-7E
|
150mV
|
20mA
|
825mV
|
DCS 300-3.5E
|
300mV
|
15mA
|
1650mV
|
DCS 600-1.7E
|
600mV
|
6.8mA
|
3300mV
|
Output
|
DCS Series 1 kW
|
Output Power
|
Voltage VDC
|
Current ADC @ 50°C
|
DCS 8-125E
|
0-8
|
0-125
|
DCS 10-100E
|
0-10
|
0-100
|
DCS 20-50E
|
0-20
|
0-50
|
DCS 33-33E
|
0-33
|
0-33
|
DCS 40-25E
|
0-40
|
0-25
|
DCS 50-20E
|
0-50
|
0-20
|
DCS 60-18E
|
0-60
|
0-18
|
DCS 80-13E
|
0-80
|
0-13
|
DCS 100-10E
|
0-100
|
0-10
|
DCS 150-7E
|
0-150
|
0-7
|
DCS 300-3.5E
|
0-300
|
0-3.5
|
DCS 600-1.7E
|
0-600
|
0-1.7
|
Output
|
DCS Series 1 kW
|
Constant Voltage Mode*
|
Ripple** (rms) mV
|
Noise** (p-p) mV
|
Transient Response time μs (typ)
|
DCS 8-125E
|
4
|
60
|
500
|
DCS 10-100E
|
4
|
60
|
500
|
DCS 20-50E
|
4
|
60
|
500
|
DCS 33-33E
|
4
|
60
|
500
|
DCS 40-25E
|
4
|
60
|
500
|
DCS 50-20E
|
4
|
60
|
500
|
DCS 60-18E
|
4
|
60
|
500
|
DCS 80-13E
|
4
|
60
|
500
|
DCS 100-10E
|
4
|
60
|
500
|
DCS 150-7E
|
12
|
160
|
500
|
DCS 300-3.5E
|
20
|
200
|
500
|
DCS 600-1.7E
|
50
|
300
|
500
|
Output
|
DCS Series 1 kW
|
Combined Regulation Line and Load %
|
Temp Co-efficient Voltage %
|
Voltage Drift % Vmax (typ)
|
DCS 8-125E
|
0.2
|
0.02
|
0.05
|
DCS 10-100E
|
0.2
|
0.02
|
0.05
|
DCS 20-50E
|
0.2
|
0.02
|
0.05
|
DCS 33-33E
|
0.2
|
0.02
|
0.05
|
DCS 40-25E
|
0.2
|
0.02
|
0.05
|
DCS 50-20E
|
0.2
|
0.02
|
0.05
|
DCS 60-18E
|
0.2
|
0.02
|
0.05
|
DCS 80-13E
|
0.2
|
0.02
|
0.05
|
DCS 100-10E
|
0.2
|
0.02
|
0.05
|
DCS 150-7E
|
0.2
|
0.02
|
0.05
|
DCS 300-3.5E
|
0.2
|
0.02
|
0.05
|
DCS 600-1.7E
|
0.2
|
0.02
|
0.05
|
Output
|
DCS Series 1 kW
|
Constant Current Mode*
|
Regulation line and load % combined
|
Ripple (rms)** mA
|
DCS 8-125E
|
0.2
|
160
|
DCS 10-100E
|
0.2
|
128
|
DCS 20-50E
|
0.2
|
25
|
DCS 33-33E
|
0.2
|
10
|
DCS 40-25E
|
0.2
|
7
|
DCS 50-20E
|
0.2
|
7
|
DCS 60-18E
|
0.2
|
6
|
DCS 80-13E
|
0.2
|
4
|
DCS 100-10E
|
0.2
|
3
|
DCS 150-7E
|
0.2
|
2
|
DCS 300-3.5E
|
0.2
|
1
|
DCS 600-1.7E
|
0.2
|
1
|
Output
|
DCS Series 1 kW
|
Input Current, A Nominal
|
Efficiency % (Typ.)
|
230V Single Phase
|
208V Three Phase
|
DCS 8-125E
|
8
|
N/A
|
82
|
DCS 10-100E
|
8
|
N/A
|
82
|
DCS 20-50E
|
8
|
N/A
|
82
|
DCS 33-33E
|
8
|
N/A
|
84
|
DCS 40-25E
|
8
|
N/A
|
84
|
DCS 50-20E
|
8
|
N/A
|
84
|
DCS 60-18E
|
8
|
N/A
|
86
|
DCS 80-13E
|
8
|
N/A
|
86
|
DCS 100-10E
|
8
|
N/A
|
86
|
DCS 150-7E
|
8
|
N/A
|
86
|
DCS 300-3.5E
|
8
|
N/A
|
86
|
DCS 600-1.7E
|
9.5
|
N/A
|
86
|