PSURGE 30.2 Mainframe
|
Charging voltage range
|
3...33kV
|
Capacitors
|
9uF / 36uF
|
Stored energy
|
5kJ at max. charging voltage
|
HV output
|
terminals inside the test cabinet
|
Power mains
|
208, 230 or 240V
50/60Hz, 1800VA
|
HV switch
|
triggered spark gap
|
Impulse polarity
|
positive, negative, alternating
|
Impulse circuit
|
floating, max. voltage between common and ground 500V AC
|
External synchronization
|
0°....360° in 1° steps
|
Dimensions (W x D x H)
|
PSURGE 30.2
|
60 x 100 x 190 cm
|
Test cabinet
|
39 x 29 x 32 cm
|
Weight
|
approx. 320 kg
|
Impulse Module PS30-CW
|
Impulse type
|
Combination wave
|
Voltage front time
|
1.2us ±30%
|
Voltage decay time
|
50us ±20%
|
Voltage undershoot
|
≤30%
|
Voltage amplitude range
|
3...30kV
|
Impulse repetition
|
Up to 6 impulses per minute at max. voltage
|
Output impedance
|
2 Ohm
|
Current front time
|
8us ±20%
|
Current decay time
|
20us ±20%
|
Current undershoot
|
≤30%
|
Current amplitude range
|
1.5...15kA
|
Impulse Module PS30-8x20
|
Impulse type
|
Current
|
Voltage front time
|
not defined
|
Voltage decay time
|
not defined
|
Undershoot
|
not defined
|
Charging voltage range
|
3...15kV
|
Impulse repetition
|
Up to 2 impulses per minute at max. current
|
Output impedance
|
0.5 Ohm
|
Current front time
|
8us ±20%
|
Current decay time
|
20us ±20%
|
Undershoot
|
≤20%
|
Current amplitude range
|
6...30kA
|
Impulse Module PS30-10x350
|
Impulse type
|
current
|
Voltage front time
|
not defined
|
Voltage decay time
|
not defined
|
Charging voltage range
|
3...15kV
|
Impulse repetition
|
Up to 2 impulses per minute at max. current
|
Output impedance
|
12 Ohm
|
Current front time
|
10us ±20%
|
Current decay time
|
350us ±20%
|
Current amplitude range
|
240...1200A
|
Impulse Module PS30-10x1000
|
Impulse type
|
Current
|
Voltage front time
|
not defined
|
Voltage decay time
|
not defined
|
Charging voltage range
|
3...15kV
|
Impulse repetition
|
Up to 2 impulses per minute at max. current
|
Output impedance
|
37 Ohm
|
Current front time
|
10us ±20%
|
Current decay time
|
1000us ±20%
|
Current amplitude range
|
80...400A
|
Coupling and decoupling network FP-SURGE 3010
|
Impulse shape
|
Combination wave only
|
EUT line voltage
|
480V AC
110V DC
|
EUT current
|
10A
|
Imp. amplitude
|
30kV / 15kA
|
Coupling modes
|
Line to neutral
Line or neutral to ground
Line and neutral to ground
|
Coupling capacitors
|
2x9uF / 1x18uF
|
Residual voltage at EUT line input
|
15% of the applied test voltage, limited to approx. 1200V
|
Dimensions (W x D x H)
|
FP-SURGE 3010
|
60 x 100 x 190 cm
|
Test cabinet
|
40 x 40 x 80 cm
|
Weight
|
FP-SURGE 3010
|
approx. 250 kg
|