»  Categories   »      Voltage Surge Generator   »    Surge Generator   »    Ring Wave Generator   »    Pulse Generator   »    EFT/Burst Generator   »    Generator  

image enlarge

Haefely PIM 110 0.5us - 100 kHz Ring Wave Impulse Module

Description

IEC 61000-4-12 and ANSI C62.41 describe a Ring Wave generator for application on power lines in a protected environment. ANSI C62.41 specifies the ring wave for testing products situated in Location Category A and B. The standard differentiates between the two categories by detailing different effective impedances of 30 and 12 ohms respectively.

IEC 61000-4-12 applies the same criteria as ANSI but adds 200 ohm source impedance and includes the testing of input/output ports.

Safety is paramount with surge impulses. The PIM 110 High Voltage connections are all linked into the System Safety Concept which prevents high voltage plugs or cables from exposing operators to potentially lethal voltages. This concept is extended to Coupling/Decoupling Networks (CDNs) connected to the PIM 110 output.

PIM 110 as part of the PSURGE 8000 Surge Platform System can be used together with Coupling/Decoupling Networks (CDNs) for superimposition of the impulse on AC and DC power lines. Single and three phase CDNs have multiplexed inputs, allowing up to 3 impulse modules to be connected. A full suite of test impulses can then be performed completely automatically and without the need to power down and reconnect the EUT.

PSURGE 8000 Surge Platform can perform all the programming functions required to perform IEC and ANSI testing without the need of a control computer. Voltage and Current monitor circuits enable quick and easy verification of the impulse shape and amplitude. In addition, peak measurement circuits are used to help determine EUT PASS/FAIL criterion.

The s integration in the WinFEAT&R control and reporting software package enhances an efficient set-up and operation of this test system. Most importantly, the test load can be transferred to a computer freeing valuable resources.

Features

  • Floating high voltage output
  • Ring wave 0.5us - 100 kHz (voltage & current)
  • 7.8kV impulse voltage
  • 12W, 30W and 200W source impedance
  • Impulse voltage & current monitors
  • Accurate phase angle synchronization

Applications

  • Single & Three phase power line systems
  • IEC 61000-4-12 Power & Data lines
  • ANSI C62.41 Power lines
  • Many IEC & EN Product standards

Other international requirements for Ring wave impulses

Specifications

Impulse Voltage

0.20 – 7.8kV ±10%

Source Impedance

12, 30 & 200 Ohms

Impulse Polarity

Positive and Negative

Repetition @ Umax

5 seconds / 12 per minute

Line Synchronization

1° steps ±1° @50Hz

Impulse rise time

V= 0.5us ±20% & I= < 1.0us

Impulse frequency

100kHz ± 10% ( V & I)

Voltage Monitor

BNC, (750/1)

Current Monitor

BNC, (12 & 30W 80A/V) 200Ω 5A/V

Impulse Output

HV connector with safety circuit

Dimensions (W x H x D)

45 x 17 x 57 cm

Weight

44 lbs (25 kg)

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E