IEC and EN standards cover testing of 3 phase AC and DC power ports. They include recommendations for the coupling and decoupling component values within the CDN. These values are largely based on the European model for AC power lines. PCD 130 includes all the IEC and EN requirements with automatic coupling path switching. PCD 130 EUT output has specially designed 4mm connectors, which provide enhanced personnel safety in relation to the high voltage impulse.
ANSI / IEEE standards are the reference for power line tests in many American standards such as TIA-968-A (FCC part 68) and Telcordia (Bellcore) GR-1089-CORE. The ANSI standard contains much the same information as the IEC but based around the American experience with AC power lines.
Impedance of the low voltage mains supply to earth is simulated by the addition of a 10 ohm resistor for IEC tests. ANSI has NO series resistor in the impulse path this difference comes from the practice in Europe of connecting ground to neutral at the distribution transformer, not the power service entry as in the USA. PCD 130 automatically sets the correct coupling elements for either IEC or ANSI as selected by the user.
The integration in the WinFEAT&R control and reporting software package enhances an efficient set-up and operation of this test system. Most importantly, the test load can be transferred to a computer freeing valuable resources.
- Three high voltage inputs
- Combination wave 1.2/50us - 8/20us
- Ring wave 100kHz
- 8kV impulse voltage
- Line voltage 690Vac phase-phase
- 32A EUT Current per phase
- Phase angle synchronization for each path
Applications
- Single & Three phase power line systems
- IEC 61000-4-5 Edition 1 & 2 Power line testing
- IEC 61000-4-12 Power line testing
- ANSI C62.41 & C62.45 Power lines
- Many IEC & EN Product standards
Maximum Impulse Voltage
|
8kV
|
Maximum AC Voltage
|
690VRMS (Phase – Phase)
|
Maximum DC Voltage
|
110V
|
Maximum AC Current
|
32ARMS
|
Maximum DC Current
|
32A
|
Residual voltage at Test supply input
|
Max. 15% of the applied impulse voltage
|
EUT Connections
|
4mm HV Safety sockets
|
Phase Sync.
|
Follows coupling path
|
Phase Sync accuracy
|
±1°
|
Coupling Elements and paths
|
Controlled by the PSURGE 8000
|
Voltage drop due to the decoupling inductors
|
≤10% with max. current and cos φ ³0.7
|
Dimensions (W x H x D)
|
45 x 50 x 57 cm
|
Weights
|
69 kg
|
Standards
CISPR 11 |
CISPR 14 - 1 |
CISPR 14 - 2 |
CISPR 14 - 2 |
CISPR 16 - 1 |
CISPR 16 - 2 |
CISPR 16 - 3 |
CISPR 16 - 4 |
CISPR 22 |
CISPR 24 |
IEC/TR EN 61000 - 1 - 1 |
IEC/TR EN 61000 - 2 - 1 |
IEC/TR EN 61000 - 2 - 3 |
IEC EN 61000 - 3 - 2 |
IEC EN 61000 - 3 - 4 |
IEC/TC EN 61000 - 3 - 5 |
IEC EN 61000 - 4 - 2 |
IEC EN 61000 - 4 - 3 |
IEC EN 61000 - 4 - 4 |
IEC EN 61000 - 4 - 5 |
IEC EN 61000 - 4 - 6 |
IEC EN 61000 - 4 - 7 |
IEC EN 61000 - 4 - 8 |
IEC EN 61000 - 4 - 9 |
IEC EN 61000 - 4 - 11 |
EN 50 081 part 1 |
EN 50 081 part 2 |
EN 55 011 |
EN 55 013 |
EN 55 014 |
EN 55 015 |
EN 55 020 |
EN 55 022 |
EN 55 024 |
EN 50 082 part 1 |
EN 50 082 part 2 |
EN 50 093 |
FCC Part 15 |
MIL-STD - 461E |
|