»  Categories   »      Biconical Antenna   »    Antenna  

image enlarge

Com-Power AB-900 Biconical Antenna 30 MHz – 300 MHz

Description

The AB-900 is a linearly polarized broadband Biconical antenna specifically designed for EMC testing. It has a frequency range of 30- 300 MHz. This antenna can be used for both emissions and immunity testing.

For immunity testing, the AB-900 antenna has a balun that is capable of accepting up to 50 Watts at its input terminals to generate electromagnetic fields.

These antennas are individually calibrated using procedures described in ANSI 63.4. The calibration data is shipped with the antenna to maximize measurement accuracy.

Both antenna models have the same mounting base, with a 1/4 inch x 20 threaded hole. It can be mounted on a Com-Power AT-100 tripod or any other antenna tripod with matching threads.

Features

  • Broadband - 30 MHz to 300 MHz
  • Transmit and receive capability
  • Individual calibration

Com-Power AB-900 Applications

Biconical antennas are used for emissions and immunity testing to meet various EMC standards specified by FCC, CISPR and EN. The broadband characteristics of the Biconical antenna make it a good choice for making sweep measurements and for automated measurement systems.

Normally, tuned dipole antennas are used for EMC site attenuation measurements for better accuracy. However, the Biconical antenna is easier to use for vertical site attenuation measurements, because of the long dipole element lengths at lower frequencies (5 meters at 30 MHz). According to ANSI 63.4 specifications, a calibrated Biconical and a log periodic antenna can be used for site attenuation measurements.

The calibration data provided with each antenna is used to calculate fi eld strength measured for the selected frequency. The antenna factor (dB/m) for the selected frequency is added to the measured output (dBV) displayed by the EMI meter to obtain field strength (dBV/m).

Specifications

Frequency Range

30 MHz - 300 MHz

Polarization

Linear

Power handling

50 Watts continuous

Impedance

50 Ω

Connector

BNC (f)

Weight

7 lbs. (3.5 kg) max.

Size (L x W)

28.75" x 52.75" (73 cm x 134 cm)

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E