Description

Steady State Food & Drug Stability Chambers

Blue M’s Steady State chambers are available in four convenient sizes. With workspace sizes of 8, 16, 32 and 58 cubic feet and a variety of temperature and humidity combinations, these chambers provide the flexibility for a wide range of uses. Watlow F4 Controller allows remote and unattended operation in setting up programs and setpoints. The F4 also allows easy control over both temperature and humidity settings in one controller. Common design elements of the chambers allow shelves to be interchanged between different models.

Features

  • Electric Incoloy-sheathed tubular heaters
  • Single stage reciprocating type compressor with Non-CFC refrigerant
  • Vapor-Flo II Humidity Generator
  • Centrifugal type blower wheel with external mounted motor for air circulation
  • Door Gasket Trough and Chamber Drain
  • Non-ramping, manual mode
  • Direct printing to a serial printer
  • Direct connection to a PLC or computer workstation
  • High definition, LCD display with four digital input lines
  • Solid Shelves included

Performance

  • Heating/Cooling Range: 0° C to + 99° C, ± 0.3° after stabilization
  • Heating/Cooling/Humidity Range: 20% to 96% RH, ±2% for relative humidity
  • Humidity Range with Optional Air Dryer: 5% RH at temperatures to +20° C

Controller – Watlow F4

  • -2 Model controlled by a single-channel type
  • -4 Model controlled by a two-channel type
  • Time/ temperature profiling controller that demands ramp and soak applications
  • Programming capability of up to 256 steps in as many as 40 profiles.
  • 16-bit microprocessor
  • Ability to function as a static set point with remote control
  • 100 Ohm Platinum RTD sensor for temperature measurement
  • Dry capacitance type humidity sensor for humidity measurement
  • EIA-232 and EIA-485 serial communications with Modbus RTU protocol

Programmable Logic Relay (PLR)

  • Teco Model SG2-20KR provides enhanced system control, replacing individual mechanical relays
  • Up to 200 lines available
  • 120 VAC input power
  • 12 AC discrete inputs
  • 8 relay outputs rated 8 amps resistive, 4 amps inductive
  • Real-time clock
  • 15 timers and 15 counters

Specifications

Model

Non-humidified

CEO 908-2

 

CEO 916-2

 

CEO 932-2

 

CEO 958-2

 

Humidified

 

CEO 908-4

 

CEO 916-4

 

CEO 932-4

 

CEO 958-4

Work Space inches/cm

W

33/84

33/84

33/84

60/152

D

28/71

28/71

28/71

28/71

H

15/38

30/76

60/152

60/152

Exterior inches/cm

W

39/99

39/99

39/99

70/178

D

32/81

32/81

32/81

32/81

H

46/117

61/155

91/231

91/231

No. of  Solid Shelves (included)

1

2

4

8

No. of Shelf Guides

4

9

19

38

Temperature Range

Low

0

High

99°C

Humidity capability

20% to 96% RH on -4 models. Based on ambient conditions of 22°C and 50% RH, limited by a 5°C dewpoint temperature.

Utilities

Refrigeration (HP)

1/2

3/4

Heater Capacity

2 kW

6 kW

Humidifier

KW

1.5

3

GPH

0.5

1

AMPS @ 208/230, 3-WIRE, 1Ph-60Hz

20

24

20

24

20

24

40

55

AMPS Breaker
Overcurrent protection recommended

30

30

30

30

30

30

50

70

Unit Weight

LBS

375

400

425

450

524

550

1850

1900

Note: Casters add 2.5" to exterior height
Control tolerance: ±0.3°C and ±2% after stabilization.
All specifications are subject to change without notice.

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E