»  Categories   »      EMC / EMI   »    EMC Analyzer   »    Analyzer  

image enlarge

Amplifier Research DER2018 Digital EMI Receiver 20 Hz – 18 GHz

Description

Amplifier Research DER2018 is a state-of-the-art DSP-based emissions receiver with an integrated computer capable of reducing your valuable test time from days to minutes. It is extremely easy to operate with all functions menu driven and displayed on a supplied 23" widescreen monitor.

This digital receiver offers continuous coverage from 20 Hz to 18 GHz with an instantaneous bandwidth of 140 MHz. Peak, Quasi-peak, Average, and RMS-Average detections are processed simultaneously up to 8,192 frequency points and interpolated with a proprietary algorithm to allow scans in seconds, catch short duration transient disturbances, and identify emissions using a fast time-base 3-D display. This receiver combines state-of-the-art sensitivity, dynamic range, accuracy and convenience of operation. It complies with CISPR-16-1-1 edition 3.0 2010-01.

The Multistar Precision DSP Receiver system includes a built-in computer and interfaces with standard data storage, high resolution video devices and features a removable hard drive. A 23" widescreen monitor, keyboard and mouse are included.

Features

Emission Testing Solutions to the following standards:

  • MIL-STD-461D, E & F
  • DO160D, E & F
  • CISPR 11/EN 55011
  • CISPR 22/EN 55022
  • CISPR 14/EN 55014
  • FCC Part 15
  • 140 MHz-wide, pre-selected, instantaneous bandwidth*

PEAK, QUASI-PEAK, AVERAGE, and RMS-AVERAGE detections are processed simultaneously at 8,192 frequency points and interpolated using a proprietary algorithm. These features enable the user to:

  • Display and record detector results as continuous spectra with 1 Hz resolution
  • Scan 4 GHz every second with the PEAK detector, i.e. covering 1 - 18 GHz in 4.25 seconds
  • Sweep 9 kHz - 30 MHz (bands A & B) in 2 seconds with all CISPR detectors
  • Process 30 - 1000 MHz (bands C & D) in 7 seconds with all CISPR detectors
  • Reduce multi-day tasks to minutes
  • Catch short-duration transient disturbances
  • Identify emissions using fast time-base 3-D display

Easy to use – all functions are menu driven.

Internal wide band calibration source expedites periodic checking and re-calibration of the receiver’s amplitude response.

Capability for user to set up, and save for future use, all of the needed test parameters including limit lines, start/stop frequencies, IF bandwidth, samples per bandwidth, dwell time at each frequency, input port selection vs. test frequency, transducer correction table, input attenuation, units to be used for the displayed level units, and more. 

*140MHz instantaneous bandwidth is available in bands C, D and E with -6dB resolution bandwidth ≥ 50 kHz. The entire bands A and B are covered instantaneously with -6dB bandwidths at least 100Hz and 9 kHz respectively. With narrower resolution bandwidth settings, the instantaneous bandwidth is proportionally reduced.

Specifications

All references to CISPR specification are to CISPR-16-1-1 edition 3.0 2010-01

All references to MIL-STD specification are to MIL-STD-461 D, E & F

Definitions: ADNL = Average displayed noise level, PDNL = Peak displayed noise level

PARAMETER

SPECIFICATION

FREQUENCY RANGE

DER2018 Base System

20 Hz–18 GHz

With CFE1840 antenna mountable down-converter. 20 Hz–40 GHz (See CFE1840 spec sheet)

MODES OF OPERATION: Spectrum Analyzer Modes

Free running

Single sweep

Video, software and external trigger

MODES OF OPERATION: Time Domain Analyzer Modes

Single Frequency

Single instantaneous sub-band

Free running

Single shot

Video, software and external trigger

FREQUENCY RESOLUTION (Display & Markers)

1 Hz

DIGITALLY PROCESSED IF FILTERS

GAUSSIAN-SHAPED, -6dB or -3dB Bandwidths selectable

20 Hz–30 MHz

Any bandwidth in the range 10 Hz -350 kHz

30 MHz–18 GHz

Any bandwidth in the range 50 Hz -1.7 MHz

18 GHz–40 GHz (with CFE1840 down-converter)

Any bandwidth in the range 50 Hz -1.7 MHz

LEVEL MEASUREMENT UNCERTAINTY

±1.0 dB (95% uncertainty interval)

STABILITY OF INTERNAL FREQUENCY STANDARD

Initial setting

±0.2 ppm

Over operating temperature range

±0.4 ppm

First year

±0.5 ppm

First ten years

±2.0 ppm

SENSITIVITY & DYNAMIC RANGE (0dB input attenuation, -6dB resolution bandwidths, Preamp OFF)

Frequency Range

(MHz)

Resolution

Bandwidth (kHz)

ADNL (dBm) (typical)

PDNL (dBm)

(max., incl. spurious)

Typical

Overload Range

(dBm)

30 Hz–1 kHz

0.01

-110

-92

-1 to +2

10 kHz–30 MHz

0.01

-140

-92

-1 to +2

1 kHz–10 kHz

0.1

Linear slope:

-90 @ 1kHz

-125 @ 10kHz

Linear slope:

-85 @ 1kHz

-115 @ 10kHz

-1 to +2

9 kHz–150 kHz

0.2

-129

-100

-1 to +2

10 kHz–150 kHz

1

-122

-100

-1 to +2

150 kHz–30 MHz

9 or 10

-112

-92

-1 to +2

30–300 MHz

100 or 120

-100

-82

-1 to +8

300-1000 MHz

100 or 120

-100

-82

-7 to +2

1-6 GHz

1,000

-97

-78

-8 to +2

6-18 GHz

1,000

-90

-69

-4 to +9

SENSITIVITY & DYNAMIC RANGE (0dB input attenuation, -6dB resolution bandwidths, Preamp ON)

Frequency Range

(MHz)

Resolution

Bandwidth (kHz)

ADNL (dBm) (typical)

PDNL (dBm)

(max., incl. spurious)

Typical

Overload Range

(dBm)

30–300 MHz

100 or 120

-118

-104

-30 to -21

300-1000 MHz

100 or 120

-118

-100

-38 to -30

1-6 GHz

1,000

-118

-98

-37 to -27

6-18 GHz

1,000

-118

-93

-37 to -16

  

1dB COMPRESSION POINT

Above overload level

THIRD ORDER INTERCEPT POINT

(0 dB input attenuation, CW signals)

Typically 10dB above overload level

DETECTORS AVAILABLE IN BOTH SPECTRUM ANALYZER AND RECEIVER MODES

PK, MIL-STD Peak hold

PK, CISPR Peak hold

QP, AVG, RMS-AVG, CISPR weighting and filtering. All detectors displayed simultaneously.

DETECTORS ADDITIONALLY AVAILABLE IN TIME DOMAIN ANALYZER MODES

AM

PRESELECTION

Bands A, B

20 Hz - < 30 MHz

Band C #1

30 MHz - < 160 MHz

Band C #2

160 MHz - < 300 MHz

Band D #1

300 MHz - < 440 MHz

Band D #2

440 MHz - < 580 MHz

Band D #3

580 MHz - < 720 MHz

Band D #4

720 MHz - < 860 MHz

Band D #5

860 MHz - < 1000 MHz

Band E #1

1 GHz - < 6 GHz

Band E #2

6 GHz - 18 GHz

Band K (with CFE1840 Down-converter)

18 GHz - < 26.5 GHz

Band Ka (with CFE1840 Down-converter)

26.5 GHz - 40 GHz

IMAGE REJECTION (0 dB input attenuation)

> 95 dB, CISPR limit > 40 dB (par 4.5.3)

IF REJECTION (0 dB input attenuation)

> 95 dB, CISPR limit > 40 dB (par 4.5.2)

RF INPUTS (Selectable, 50 Ohm, unbalanced, front panel)

Regular RF input

Remote LN1G18 Pre-amp input with DC Bias

CFE1840 Down-converter input

MAX DC VOLTAGE AT ANY RF INPUT

0 VDC maximum

Standards

CISPR
CISPR 11
CISPR 14 - 1
CISPR 14 - 2
CISPR 14 - 2
CISPR 16 - 1
CISPR 16 - 2
CISPR 16 - 3
CISPR 16 - 4
CISPR 22
CISPR 24
IEC
IEC/TR EN 61000 - 1 - 1
IEC/TR EN 61000 - 2 - 1
IEC/TR EN 61000 - 2 - 3
IEC EN 61000 - 3 - 2
IEC EN 61000 - 3 - 4
IEC/TC EN 61000 - 3 - 5
IEC EN 61000 - 4 - 2
IEC EN 61000 - 4 - 3
IEC EN 61000 - 4 - 4
IEC EN 61000 - 4 - 5
IEC EN 61000 - 4 - 6
IEC EN 61000 - 4 - 7
IEC EN 61000 - 4 - 8
IEC EN 61000 - 4 - 9
IEC EN 61000 - 4 - 11
European
EN 50 081 part 1
EN 50 081 part 2
EN 55 011
EN 55 013
EN 55 014
EN 55 015
EN 55 020
EN 55 022
EN 55 024
EN 50 082 part 1
EN 50 082 part 2
EN 50 093
American
FCC Part 15
MIL-STD - 461E